Study 133148
Study 133148
Video Endoscopy cable assemblies
Very flexible cable with an extremely good mechanical behaviour.
Withstands miscellaneous sterilisation process.
- Nexans ref. : Etude 133148
- Country ref. : Etude 133148
Description
Description
Standards
-
ProductNexans specification
1- CORE (A)
AWG 36 stranded Silver plated alloy conductor.
MFA insulation.
Diameter : 0.27 +/- 0.03 mm.
CORE (B)
Filler :
AWG 44 Silver plated copper wire.
MFA insulation.
Nominal diameter : 0.18 +/- 0.03 mm.
2- CORE (C)
2 Cores (A) + 2 Cores (B)
PTFE tape
Nominal diameter : 0.70 mm
AWG 44 Silver plated copper spiral shield.
Nominal diameter : 0.80 mm
MFA Jacket
Diameter : 0.92 +/- 0.05 mm
3- CORE (D)
4 Cores (A) twisted
PTFE tape
Nominal diameter : 0.78 mm
AWG 44 Silver plated spiral shield.
Nominal diameter : 0.88 mm
MFA jacket.
Diameter : 1.00 +/- 0.05 mm
4- CORE (E)
4 Cores (A) twisted
Nominal diameter : 0.65 mm
Assembly
1 Core (C) + 1 Core (D) + 2 Cores (E) twisted.
Nominal diameter : 1.92 mm
5- PTFE Tape
Nominal diameter : 1.98 mm
6- DRAIN Wire
7 x 40 AWG Silver plated copper alloy.
7- SHIELD
AWG 40 silver plated copper alloy braid.
85% coverage.
Nominal diameter : 2.37 mm.
8- PTFE tapes
Nominal diameter : 2.65 mm
9- Medical Polyurethane Jacket
Thickness : 0.50 mm min.
Nom. Diameter : 4.05 mm.
Max. Diameter : 4.15 mm.
Internal diameter 2.85 mm to 3.05 mm
Main Characteristics
Minimum static bending radius : 5 x Overall diameter.
Characteristics
Characteristics
结构特性
结构特性
尺寸特性
尺寸特性
电气特性
电气特性
机械特性
机械特性
使用特性
使用特性